Technical Details of Cartridges for Reduced Drug-Container Interaction
Dimensions and Cosmetics
Highly accurate production techniques and decades of experience has the result of SCHOTT being able to facilitate tighter dimensions and premium surface quality that reliably ensures reduced drug-container-interaction, superior container processing, device compatibility and reliable performance at point of use.
The flange form of a cartridge is determined by the filling procedure or the crimp solution: Lined seals have a proven fit with our cartridges and are optimized for device performance. These seals are designed in accordance with industry standards for single-dose or multi-dose applications in various formats for liquid application. For bigger openings, serum stoppers for vials and their respective crimps are used.
We offer the following flanges with our SCHOTT cartridges:
We offer tight geometric tolerances:
Following options upon request:
|
Hydrolytic Resistance
Improved Type I FIOLAX® borosilicate glass – FIOLAX® CHR 'Controlled Hydrolytic Resistance' for cartridges. FIOLAX® CHR fullfills and comply with PH.Eur., USP and JP international standards.
Strength
Break-resistant cartridges made out of FIOLAX® with chemical strengthening is the customized option for the harshest environmental requests. Chemical toughening of FIOLAX® with a depth of penetration of 36 uM helps to fullfill the requested additional side-compression specification of >/= 350 lbf in mean.
Material / Glass Type
SCHOTT cartridges are produced using a choice of glass materials, each with their own individual properties for specific applications. These materials include:
- FIOLAX®
- FIOLAX® (CHR) clear
Technical details for our cartridges
Standardized quality level according to ISO
The StandardLine includes:
- Production in cGMP environment
- Statistical in-process control
- Dimensional and cosmetic AQL levels according to ISO
- 100% camera inspection of dimensional parameters
- Pen cartridges: ISO 13926-1: 2018
- Dental cartridges: ISO 11040-1: 2015
- Double Chamber cartridges (1 ml)
Customized product specifications and quality levels
TopLine options include:
- Production in cGMP environment
- Statistical in-process control
- Customized dimensional and cosmetic AQL levels
- 100% camera inspection of dimensional and cosmetic parameters
- Pen cartridges: ISO 13926-1: 2018
- Double Chamber cartridges (customized formats available)
- Break Resistant cartridges
- Ready-to-use: see cartriQ®